Important Notes
- Use only 20 mL scintillation vials (have to be special VWR brand under the instrument)
- Do not fill the scintillation vials with more that 8 mL of solvent
- Use the special green key to insert the 20 mL vial into the holder
- Do not put vials in slot “A”; it is bad and tends to drop samples
- No sign-ups; first come first serve (~ 5 min/vial)
- Do not edit prior person’s program. Wait for their run to complete and then start your run
Safety notes
- If removing TFA or a strong acid, wear appropriate PPE
- Close sash of the fume hood box when not in use
- If transporting vials between rooms, always close with the scintillation vial cap
- Do a “wash” vial at the end of your cycle with 20 mL iPrOH to clean out the tool
Procedure
- Put vials in instrument using the green key to open up the levers. Remember final wash vial. Close fume hood.
- Check list of solvents and select the appropriate method to help reduce risk of solvent bumping (choloroform=volatile)
- Log-into computer. Log-in: NanoBio
- Close out the previous person’s run by directly exiting the program
- Open program. Don’t change any settings other than “method” (see step 2)
- Add vials starting at slot A
- On list, delete vial A and any unfilled slots
- On the software, hit run.